Aerosol Neutralizer Body
BMI fabricates an Aerosol Neutralizer Body with mounting brackets for two Polonium-210 ionizing strips
(Model 2U500 from NRD LLC(TM)). The neutralizer is appropriate for imparting a Boltzmann charge distribution
to particles in sample flow rates between 1 and 5 lpm and total particle concentrations up to 50,000/cc.
Size distributions observed using the BMI neutralizer have been compared successfully against those obtained
using other commercially available neutralizers. The neutralizer is appropriate for use with differential
mobility analyzers and for diluted atomizer flows. The unit is shipped without the Po-210 strips, which
must be purchased directly by the end-user and are easily installed.
Key features of the neutralizer include:
- Compact design, 9" long by 1.25" diameter stainless steel tubing with high-vacuum brazed construction
- Easy disassembly to install new sources and to clean inside surfaces
- Easy connections to existing instrumentation using 1/4" OD tubulations
- Different models available for different particle concentrations and flow rates
- Design can be easily adapted to mount other components (heaters, vapor sources, chillers, EM grids) to expose sampled aerosol to controlled environment conditions for other applications
For pricing see our price sheet.
|
|
 |
|
|